TRACE
200
By stepping up to the Trace 200 Model line of analyzers
you get either a Class I Divsion II or Class I Division I rating. You also
have the option of upgrading the standard galvanized steel conduits
to corrosion proof stainless steel conduits.
CORROSION PROOFING
Using stainless steel instead of galvanized steel in the
construction of the wiring conduits will reduce the risk of corrosion of
the instrument; all of the conduits in the analyzer are replaced with their
stainless steel counterparts.
DIAGNOSTIC ALARM
The alarm provides contact output to alert operations in
the event of one of the following fault conditions:
- Out of tape
- Motor failure
- Optical system failure
- Auto Calibration failure
SMARTALK
SMARTalk provides complete remote operation via personal
computer or host with true bidirectional communication. The operator does
not have to travel to the measurement site to alter operation, initiate
calibration, or view data. Controls, reports, and data are available at
lab or office via a personal computer link. Remote access reduces the chance
of "no fault found" service calls where on site trips find no
fault with the equipment.
LCD GRAPH DISPLAY
The display is a paperless chart recording of the concentration
with selectable time scale to make comparisons and graphically display operating
trends over time.
CABINET HEATER
A heater with thermostat control is integrated into the
cabinet which provides freeze protection throughout the cold winter months.
This will prevent the humidifier solution from freezing in sub zero temperatures
causing the unit to malfunction. This will drop the lowest allowable operating
temperature 15°C to -10°C.
VAPORIZING REGULATOR
The vaporizing regulators vaporize a liquid into a gas
then perform a single stage pressure reduction on the gas. Pressure drops
through pressure regulators may cause cooling which allows liquid to form.
The vaporizing regulator puts heat back into the flow media to insure that
is stays a gas.
CHART RECORDERS
A hard copy record of 4-20mA output is available in:
- General Purpose / Division II
- Division I
AUTO CALIBRATION
The auto calibration provides software algorithm, contact
outputs, solenoid valves, tubing, and customer connections for automatically
calibrating analyzer at pre-selected time intervals.
REMOTE DISPLAY AND ALARM
This option uses the 4-20 output from the analyzer to generate
a digital display of concentration and adjustable alarm contact outputs
CAL TAP
The calibration tap point includes the connection fittings
and tubing together with a three-way ball valve to easily connect calibration
gas to the analyzer.
BACK-LIT DISPLAY
A new option available on the TRACE analyzer
is a light behind the LCD to make operation even more user friendly. |
 |
DISPLAY
- Alpha Numeric LCD
- 128 x 64 pixel
TEMPERATURE RANGES
- -10°C to 50°C (operating)
- 0°C to 70°C (storage)
ANALOG - Isolated 4-20mA
ANALYTICAL PERFORMANCE
- Resolution: 1 ppb
- Accuracy: ±2%
- Repeatability: ±1%
- Linearity:±1%
- Drift: Nil
- Temp. Coefficient: 0.01% / °C
- Analysis Time: 0.75 Second
- Detection Ranges:
0-1 ppm
0-50 ppm
0-500 ppm
AREA CLASSIFICATION
WEIGHT - Approx. 50 lbs; 22.7 kg
DIMENSIONS
- 16" X 18" X 10"
POWER
REQUIRED
- 110VAC, 60 Hz or
- 220VAC, 50 Hz or
- 12VDC, or
- 24VDC
ADDITIONAL FEATURES
(compared to the previous model)
OPTIONAL FEATURES
(features not optional on previous model are
bold)
- Diagnostic Alarm
- SMARTalk
- LCD Graph Display
- Auto Calibration
- Cabinet Heater
- Vaporizing Regulator
- Chart Recorders
- Cabinet Stand
- Remote Display and Alarms
- TRACErase
- Solar Power
- Multi Stream
- Sample Filter
- Back-Lit Display
- Corrosion Proofing
|
CABINET STAND
The cabinet stand is a pipe stand with mounting holes for
outdoor and indoor use.
TRACERASE
TRACErase uses technology based on a catalytic combustion
process to oxidize vented samples thus preventing fugitive emissions from
entering the atmosphere without causing back pressure on the analyser. Packaged
to provide explosion proof protection, this unit may be used in hazardous
locations.
SOLAR POWER
This option includes batteries, a solar panel, and charge
control to maintain analyzer operation in remote locations
MULTI-STREAM
This option allows the analyzer to select and analyze more
than one sample source. This is ideal for any situation where more than
one sample source is present. |