FEATURES

Interference Free Detection

Digital Electronics

ASTM Method References

Temperature Compensation

Operator Keypad

LED Light Source

Push-Button Calibration

Dual Processor Power

18-Bit A/D Conversion

Status Report Log

Graphic Data Displays

Swept Volume Humidifier

4-20 Output


EASE OF INSTALLATION
AND
OPERATION

Compact Size

Hinged-door Access

Light Weight

Word Menu Displays

Flow-through Filter

Latch Block open to
Change Tape


H2S Detection 
TRACE 400

High Range


TRACE 400

When dealing with H2S amounts in excess of 500ppm, it is necessary to add a dilution system to decrease the concentration to measurable levels by the Trace detector. While this decrease the Resolution slightly, it is necessary for detection at these levels.

AIR DEMAND (CLAUS PLANT)

Determines the H2S content for total air required to combust H2S in the feed gas to the Claus Plant. Two thirds of the H2S is subtracted from the total air demand to yield a control signal to the acid gas/air controller to maintain the correct stoichiometric ratio of H2S/SO2 for maximum sulfur recovery. It provides a fast acting feed-forward control of the burner air.

DIAGNOSTIC ALARM

The alarm provides contact output to alert operations in the event of one of the following fault conditions:

  1. Out of tape
  2. Motor failure
  3. Optical system failure
  4. Auto Calibration failure

SMARTALK

SMARTalk provides complete remote operation via personal computer or host with true bidirectional communication. The operator does not have to travel to the measurement site to alter operation, initiate calibration, or view data. Controls, reports, and data are available at lab or office via a personal computer link. Remote access reduces the chance of "no fault found" service calls where on site trips find no fault with the equipment.

LCD GRAPH DISPLAY

The display is a paperless chart recording of the concentration with selectable time scale to make comparisons and graphically display operating trends over time.

CABINET HEATER

A heater with thermostat control is integrated into the cabinet which provides freeze protection throughout the cold winter months. This will prevent the humidifier solution from freezing in sub zero temperatures causing the unit to malfunction. This will drop the lowest allowable operating temperature 15°C to -10°C.

VAPORIZING REGULATOR

The vaporizing regulators vaporize a liquid into a gas then perform a single stage pressure reduction on the gas. Pressure drops through pressure regulators may cause cooling which allows liquid to form. The vaporizing regulator puts heat back into the flow media to insure that is stays a gas.

CHART RECORDERS

A hard copy record of 4-20mA output is available in:

  • General Purpose / Division II
  • Division I

AUTO CALIBRATION

The auto calibration provides software algorithm, contact outputs, solenoid valves, tubing, and customer connections for automatically calibrating analyzer at pre-selected time intervals.

REMOTE DISPLAY AND ALARM

This option uses the 4-20 output from the analyzer to generate a digital display of concentration and adjustable alarm contact outputs

MULTI-STREAM

This option allows the analyzer to select and analyze more than one sample source. This is ideal for any situation where more than one sample source is present.

CAL TAP

The calibration tap point includes the connection fittings and tubing together with a three-way ball valve to easily connect calibration gas to the analyzer.

DISPLAY

  • Alpha Numeric LCD
  • 128 x 64 pixel

TEMPERATURE RANGES

  • -10°C to 50°C (operating)
  • 0°C to 70°C (storage)

ANALOG - Isolated 4-20mA

ANALYTICAL PERFORMANCE

  • Resolution: 10 ppb
  • Accuracy: ±2%
  • Repeatability: ±1%
  • Linearity:±1%
  • Drift: Nil
  • Temp. Coefficient: 0.01% / °C
  • Analysis Time: 0.75 Second
  • Detection Ranges:
     0-100 %

AREA CLASSIFICATION

  • Division II
  • Division I

WEIGHT - Approx. 75 lbs; 34 kg

DIMENSIONS - 30" X 30" X 12"

POWER REQUIRED

  • 110VAC, 60 Hz or
  • 220VAC, 50 Hz or
  • 12VDC, or
  • 24VDC

ADDITIONAL FEATURES
(compared to the previous model)

  • Air Demand (Claus Plant)

OPTIONAL FEATURES
(features not optional on previous model are bold)

  • Diagnostic Alarm
  • SMARTalk
  • LCD Graph Display
  • Auto Calibration
  • Cabinet Heater
  • Vaporizing Regulator
  • Chart Recorders
  • Cabinet Stand
  • Remote Display and Alarms
  • TRACErase
  • Solar Power
  • Multi Stream
  • Sample Filter
  • Back-Lit Display
  • Corrosion Proofing

CABINET STAND

The cabinet stand is a pipe stand with mounting holes for outdoor and indoor use.

TRACERASE

TRACErase uses technology based on a catalytic combustion process to oxidize vented samples thus preventing fugitive emissions from entering the atmosphere without causing back pressure on the analyser. Packaged to provide explosion proof protection, this unit may be used in hazardous locations.

SOLAR POWER

This option includes batteries, a solar panel, and charge control to maintain analyzer operation in remote locations

CORROSION PROOFING

Using stainless steel instead of galvanized steel in the construction of the wiring conduits will reduce the risk of corrosion of the instrument; all of the conduits in the analyzer are replaced with their stainless steel counterparts.

BACK-LIT DISPLAY
A new option available on the TRACE analyzer is a light behind the LCD to make operation even more user friendly.