Trace Model 100: H2S Detection

This line of analyzers adds much more functionality and customizability compared to the 050 model, yet retains similar portability features. View the additional and optional features in the specifications box to see all of the enhancements.

DILUTION FILTER

On the 100 model line of analyzers it is possible to achieve a 0-100% detection range with a specially designed dilution filter. This is currently the only available (other dilution methods are available on 400 models and higher) model with this dilution system as an option.

PRESSURE REGULATOR

This device is used to reduce a high pressure, supply pressure, or inlet pressure in one stage to a lower pressure. The pressure reducing regulator is designed to meet the stringent requirements of a process analyzer sample conditioning system. The low internal volume reduces lag time thus more nearly approaching real-time analysis.

SAMPLE FILTER

This device is used to remove unwanted contamination in a system. Bypass filters allow a continuous sweeping action across the surface of the filter element. This sweeping action prolongs the life of the filter element and also insures the sample that passes through the element is representative of the current system condition.

CAL TAP

The calibration tap point includes the connection fittings and tubing together with a three-way ball valve to easily connect calibration gas to the analyzer.

DIAGNOSTIC ALARM

The alarm provides contact output to alert operations in the event of one of the following fault conditions:

  1. Out of tape
  2. Motor failure
  3. Optical system failure
  4. Auto Calibration failure

SMARTALK

SMARTalk provides complete remote operation via personal computer or host with true bidirectional communication. The operator does not have to travel to the measurement site to alter operation, initiate calibration, or view data. Controls, reports, and data are available at lab or office via a personal computer link. Remote access reduces the chance of "no fault found" service calls where on site trips find no fault with the equipment.

LCD GRAPH DISPLAY

The display is a paperless chart recording of the concentration with selectable time scale to make comparisons and graphically display operating trends over time.

SAMPLE PUMP

A miniature DC powered rotary vein pump to provide sample flow from atmospheric sources. This option is necessary if there is no line pressure.

CABINET HEATER

A heater with thermostat control is integrated into the cabinet which provides freeze protection throughout the cold winter months. This will prevent the humidifier solution from freezing in sub zero temperatures causing the unit to malfunction. This will drop the lowest allowable operating temperature 15°C to -10°C.

VAPORIZING REGULATOR

The vaporizing regulators vaporize a liquid into a gas then perform a single stage pressure reduction on the gas. Pressure drops through pressure regulators may cause cooling which allows liquid to form. The vaporizing regulator puts heat back into the flow media to insure that is stays a gas.

CHART RECORDERS

A hard copy record of 4-20 mA output is available in:

  • General Purpose / Division II
  • Division I

AUTO CALIBRATION

The auto calibration provides software algorithm, contact outputs, solenoid valves, tubing, and customer connections for automatically calibrating analyzer at preselected time intervals.

BATTERY POWER

The Trace 100 model line of analyzers can be supplied with an optional battery power supply to complete remote operation. This feature is available on the 050 model line also!

CABINET STAND

The cabinet stand is a pipe stand with mounting holes for outdoor and indoor use.

TRACErase

TRACErase uses technology based on a catalytic combustion process to oxidize vented samples thus preventing fugitive emissions from entering the atmosphere without causing back pressure on the analyzer. Packaged to provide explosion proof protection, this unit may be used in hazardous locations.

SOLAR POWER

This option includes batteries, a solar panel, and charge control to maintain analyzer operation in remote locations

REMOTE DISPLAY AND ALARM

This option uses the 4-20 output from the analyzer to generate a digital display of concentration and adjustable alarm contact outputs

MULTI-STREAM

This option allows the analyzer to select and analyze more than one sample source. This is ideal for any situation where more than one sample source is present.

BACK-LIT DISPLAY

A new option available on the TRACE analyzer is a light behind the LCD to make operation even more user friendly.

TRACE 100


Sample Conditioning

Specifications

DISPLAY
  • Alpha Numeric LCD
  • 128 x 64 pixel
TEMPERATURE RANGES
  • -10°C to 50°C (operating)
  • 0°C to 70°C (storage)
ANALOG
  • Isolated 4-20mA
ANALYTICAL PERFORMANCE
  • Resolution: 1 ppb
  • Accuracy: ±2%
  • Repeatability: ±1%
  • Linearity:±1%
  • Drift: Nil
  • Temp. Coefficient: 0.01% / °C
  • Analysis Time: 0.75 Second
DETECTION RANGES
  • 0-1 ppm
  • 0-50 ppm
  • 0-500 ppm
AREA CLASSIFICATION
  • General Purpose

WEIGHT

  • Approx. 20 lbs; 9 kg
DIMENSIONS
  • 14" X 16" X 8"
POWER OPTIONS
  • 110VAC, 60 Hz
  • 220VAC, 50 Hz
  • 12VDC
  • 24VDC
ADDITIONAL FEATURES

(vs lower numbered models)

  • Pressure Regulator
  • Cal Tap
OPTIONAL FEATURES
  • Battery Power
  • SMARTalk
  • LCD Graph Display
  • Back-Lit Display
  • Not available on lower models:

  • Sample Filter
  • Dilution Filter
  • Diagnostic Alarm
  • Auto Calibration
  • Sample Pump
  • Cabinet Heater
  • Vaporizing Regulator
  • Chart Recorders
  • Cabinet Stand
  • Remote Display and Alarms
  • TRACErase
  • Solar Power
  • Multi-Stream